Safety of boneless reconstruction of the mandible with a CAD/CAM designed titanium device: The replica cohort study.

2021 
Abstract Objective We evaluated the safety of REPLICA, a CAD/CAM-designed patient-specific titanium mandible, in patients with mandibular defects not suitable for reconstruction with traditional techniques. Patients and methods We performed a cohort study with a composite primary outcome assigned at the end of a 1-year follow-up. The outcome was assigned in the presence of all the following: 1) absence of intraoral or skin extrusion of REPLICA; 2) decrease or cessation of oral pain; 3) stability or increase in mouth opening; 4) resumption of oral feeding without the need of nasogastric tube; 5) absence of fracture at multidetector computer tomography (MDCT); 6) absence of displacement (MDCT); 7) absence of screw loosening (MDCT). The secondary outcome was the patient-reported QOL at 6 months of follow-up as detected by the EORTC QLQ-C30 and QLQ-HN74) underwent reconstruction of mandibular defects with REPLICA at our Unit. The primary outcome was reached by 14 of the 18 patients. QOL data were available for 15 patients at the 6-month follow-up, showing a good profile of general and disease-specific QOL. Conclusion REPLICA offered a safe solution at 1-year for the treatment of mandibular defects not suitable for reconstruction with traditional techniques, and was associated with subjective well-being and satisfaction. Further studies are needed to assess the full range of indications of REPLICA.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    30
    References
    1
    Citations
    NaN
    KQI
    []