A high-precision multi-channel TAC and QAC module for the neutron detection wall*

2015 
A single width NIM module that includes eight channels of the time-to-amplitude converter (TAC) and the charge-to-amplitude converter (QAC) is introduced in the paper, which is designed for the large neutron wall detector to measure charge (energy) and time interval simultaneously. The module adopts a high precision gated integration circuit to realize TAC and QAC. The input range of TAC is from 30 ns to 1 μs, and the input range of QAC is from 40 pC to 600 pC. The linearity error of TAC is lower than 1.28%, and the time resolution of TAC is less than 0.871%. The linearity error of QAC is lower than 0.81%, and the resolution of QAC is better than 0.936%.