Comparison of two assays measuring human chorionic gonadotrophin serum concentrations in pregnancy termination and trophoblastic or non-trophoblastic tumours

2017 
AbstractBackground: Differences in human chorionic gonadotrophin (hCG) results provided by the commercial immunoassays reflect the heterogeneity of antibodies and the use of suboptimal standards. As a consequence, the principal forms of hCG and metabolites are differentially detected and the hCG tests are not suited for the same clinical applications. Conflicting results are available in the literature regarding which hCG variants are recognized by the Roche Elecsys hCG + β test. The aim of our study was to compare the hCG concentrations provided by the Siemens Immulite 2000 test and the Roche test as well as to assess the concordance between both assays.Methods: In this purpose, 152 samples obtained from women and 44 samples from men were analysed by both tests during the follow-up of pregnancy termination, gestational trophoblastic disease and malignancies. The intermediate precision of the Roche test was also investigated on a pool with a low hCG concentration.Results and conclusions: The hCG concentra...