Abstract Thermal sensitive paints (TSPs) are used for global nonintrusive detection of boundary layer transition in flow over the surface of wind tunnel research models. Since the transition is a transient process, the TSP should have a fast response characteristic. A low paint thermal conductivity is required for fast response. A thin-film thermal conductivity meter (TFTCM) was designed and built to measure thermal conductivity of the TSPs, which are typically between 50 and 150/μm thick. In this paper, the design and operating features of the TFTCM are described. Measurement of the thermal conductivity with this TFTCM of three standard thin-film low conductivity specimens, Kapton, Teflon, and Borofloat glass, showed good agreement with the manufacturer quoted values, thus validating the instrument and the procedure. Consistently repeatable values for thermal conductivity ( k =0.41±0.02 W/mK) were also obtained for the TSP specimen (TSB-B, 75 /μm) tested.