Tumor Ablation Using 3-Dimensional Electromagnetic-Guided Ultrasound Versus Standard Ultrasound in a Porcine Model

2019 
Introduction. The objective of this study was to compare the placement of ablation needles using 3-dimensional electromagnetic-guided ultrasound (guided) to standard ultrasound guidance (standard) in both laparoscopic surgery and open surgery. Endpoints for this study included targeting accuracy and number of required needle withdrawals and reorientations. Methods. Using a porcine model, fiducial markers were placed into the kidney and liver to represent tumors. Navigation and identification of target sites was achieved using standard or guided ultrasound. Intraprocedural observations as well as the number of needle placement attempts per target were recorded. Three board-certified general surgeons performed the navigation and ablation procedures. After completion of the navigation and ablation procedures, necropsy was performed. The position of the ablation zones relative to the fiducial markers was recorded. Results. A total of 48 procedures were performed across 6 animals (50% open and 50% laparoscopic...