Performance of a Factory-Calibrated Real-Time Continuous Glucose Monitoring System Utilizing an Automated Sensor Applicator

2018 
Abstract Background: This study assessed the accuracy of a factory-calibrated 10-day real-time continuous glucose monitoring (CGM) system (G6), which includes an automated sensor applicator. Methods: Seventy-six participants with insulin-treated diabetes were enrolled at four U.S. sites as part of a larger study of G6 system performance. In-clinic visits for frequent comparative blood glucose measurements using a reference instrument (YSI) were conducted on days 1, 4–5, 7, and/or 10 of system use. Accuracy evaluation included the proportion of CGM values that were within ±20% of YSI reference value for glucose levels >100 mg/dL and ±20 mg/dL for YSI glucose levels ≤100 mg/dL (%20/20), the analogous %15/15 and %30/30, and the mean absolute relative difference (MARD) between temporally matched CGM and YSI values. Participants calibrated the systems once daily. Raw sensor data were reprocessed using assigned sensor codes and a factory-calibration algorithm. Results: Reprocessed data from 62 participants (25 ...