Computational Simulation of Array-based Electroporation in the Cochlea

2018 
We present a computational model for predicting electric field distributions following array-based closed-loop electroporation in the cochlea. The model geometry was reconstructed from magnetic resonance images of the guinea pig cochlea and an eight-channel electrode array embedded within this geometry. The model’s electrode voltage output waveform was obtained from electric potential mapping conducted in physiological solution following constant-current stimulation using the electrode array. Our simulations predict that a tandem electrode configuration with four ganged cathodes and four ganged anodes produces a larger area in target tissue where the electric field is within the range for successful gene transfer compared to an alternate paired anode-cathode electrode configuration. These findings corroborate published in vivo evidence comparing the two configurations and support the utility of the developed model as a tool to optimize the efficacy of electroporation electrodes.