Accuracy of the Barrett Universal II formula integrated into a commercially available optical biometer when using a preloaded single-piece intraocular lens.

2021 
Purpose To compare the commonly used formulas for intraocular lens (IOL) selection using IOLMaster®700 (Carl Zeiss Meditec) and to evaluate the Barrett Universal II (BU-II) formula accuracy when using the Vivinex™ iSert® XY1 IOL (Hoya Corporation Medical Division). Methods A retrospective chart review was performed that included patients who underwent uneventful cataract surgery with in-the-bag insertion of Vivinex™ iSert® XY1 IOL. Prediction errors at 3 months postoperative of IOLMaster® 700 with Haigis, Holladay 1, SRK/T, and BU-II formulas were compared. As a subgroup analysis, we focused on the axial length (AL) and IOL power. AL subgroup analysis was based on the following AL subgroups: short ( 25.5 mm). IOL power subgroup analysis was based on the following IOL power subgroups: low (≤18.0 diopters [D]), medium (18.5-24.0 D), and high (≥24.5 D). Results This study included 590 eyes of 590 patients. Overall, the four IOL calculation formulas appeared to be similarly accurate. In the long AL subgroup, the BU-II formula had a significantly lower absolute error (AE) than the Holladay 1 formula. In the low-power subgroup, the BU-II formula had a significantly lower AE than the Holladay 1 and SRK/T formulas. On the other hand, in the high-power subgroup, the BU-Ⅱ formula was significantly less accurate than the SRK/T formula and also appeared to be worse than the Holladay 1 formula (P = 0.052). Conclusion The BU-II formula might be less accurate when using a Vivinex™ iSert® XY1 IOL of 24.5 D or greater.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    20
    References
    0
    Citations
    NaN
    KQI
    []