Multi-center validation of the Response Biomedical Corporation RAMP® NT-proBNP assay with comparison to the Roche Diagnostics GmbH Elecsys® proBNP assay ☆

2007 
Abstract Background NT-proBNP measurements aid in the evaluation of patients with suspected heart failure (HF) and may facilitate risk stratification in patients with HF and acute coronary syndrome (ACS). Point-of-care (POC) assays may provide more timely results and potentially improve patient outcomes. Methods We evaluated the analytical performance of the Response Biomedical Corporation whole blood RAMP amino-terminal pro-B type natriuretic peptide (NT-proBNP) POC assay compared to the Roche Elecsys proBNP (NT-proBNP) assay. Results Intra-day and total imprecision (% CV) ranged from 5.5% to 10.3% at 140, 449 and 1675 ng/L. The lowest concentration that yields a 20% CV was 57 ng/L. The lower limit of detection was 18 ng/L. The upper limit of linearity was validated to 23,428 ng/L with an average recovery of 95%. Correlation by Passing and Bablok regression yielded RAMP = 1.01 Elecsys + 14.6, r  = 0.98 ( n  = 540; range of Elecsys values  35,000). Concordance of RAMP versus Elecsys using cut-offs of 125 ng/L for subjects p Conclusions The POC Response Biomedical RAMP NT-proBNP assay provides comparable results that measured on the FDA cleared Roche Elecsys central laboratory platform.