Discordant vascular parameter measurements in diabetic and non-diabetic eyes detected by different optical coherence tomography angiography devices.

2020 
Purpose To compare quantitative changes in macular parameters in diabetic patients detected by two optical coherence tomography angiography (OCTA) instruments. Methods 80 phakic eyes were classified as no diabetes, diabetes without diabetic retinopathy (DR), mild non-proliferative diabetic retinopathy (NPDR), and severe NPDR or proliferative DR (PDR). OCTA was performed using devices from two manufacturers (Zeiss and Heidelberg). Superficial and deeper vascular skeleton density (SVSD, DVSD), superficial and deeper vessel area density (SVAD, DVAD), choriocapillaris flow voids (CCFV), and choroidal flow voids (CFV) were calculated. Inter-device comparisons were performed using the size comparison index (SCI) and the discrepancy index (DI). Results The two devices were inconsistent in SVSD, DVSD, DVAD, CCFV and CFV parameters (all P 0.05). The mean DI of CCFV was statistically different between the four groups (P < 0.001). Conclusions The two instruments were largely inconsistent in the measurement of macular parameters relevant to DR. The choice of imaging device can impact OCTA analytics and should be taken into account when drawing conclusions about DR-related changes.