Bone conducted vibration is an effective stimulus for otolith testing in cochlear implant patients.

2021 
Background Treatment with a cochlear implant (CI) poses the risk of inducing a behaviorally unmeasurable air-bone gap leading to false negative absence of cervical and ocular vestibular evoked myogenic potentials (cVEMPs, oVEMPs) to air conducted sound (ACS). Objective To investigate VEMP response rates to ACS and bone conducted vibration (BCV) in CI patients and the applicability of the B81 transducer for BCV stimulation. Methods Prospective experimental study including unilateral CI patients, measuring cVEMPs and oVEMPs to ACS and to BCV, comparing response rates, signed asymmetry ratios, latencies, and amplitudes. Results Data of 13 CI patients (mean age 44±12 years) were analyzed. For the CI side, oVEMP and cVEMP response rates were significantly higher for BCV (77%cVEMP, 62%oVEMP) compared to ACS (23%cVEMP, 8%oVEMP). For the contralateral side, no difference between response rates to ACS (85%cVEMP, 69%oVEMP) and BCV (85%cVEMP, 77%oVEMP) was observed. Substantially higher asymmetries were observed for ACS (-88±23%for cVEMPs, -96±11%for oVEMPs) compared to BCV (-12±45%for cVEMPs, 4±74%for oVEMPs). Conclusions BCV is an effective stimulus for VEMP testing in CI patients. The B81 is a feasible stimulator.