THE EFFECT OF CERAMIC FEMORAL HEAD MATERIAL COMPOSITION ON POLYETHYLENE STRUCTURE AND OXIDATION IN TOTAL HIP BEARINGS

2018 
IntroductionIn total hip arthroplasty (THA), polyethylene (PE) liner oxidation leads to material degradation and increased wear, with many strategies targeting its delay or prevention. However, the effect of femoral head material composition on PE degradation for ceramic-PE articulation is yet unknown. Therefore, using two different ceramic materials, we compared PE surface alterations occurring during a series of standard ceramic-PE articulation tests.Materials and MethodCeramic-PE THA bearings were tested in a simulator, using ASTM F2003-02, ASTM F1714-96 (2013) and ISO 14242:1–3 standards. Acetabular liners (Apex-Link PolyTM, OMNI Life Science, East Taunton, MA, USA) were articulated against O28 mm Si3N4 femoral heads (Amedica Corp., Salt Lake City, UT, USA). For comparison, ArCom® PE liners (Biomet Inc. Warsaw, IN, USA) were also tested against O28 mm zirconia-toughened alumina (ZTA) femoral heads (BIOLOX®delta, CeramTec GmbH, Plochingen, Germany), under the same conditions. After 5 million cycles of ...