Mid-term results of everolimus-eluting stent in a Japanese population compared with a US randomized cohort: SPIRIT III Japan Registry with harmonization by doing.

2012 
UNLABELLED: To address safety concerns with first-generation drug-eluting stents (DESs), the everolimus-eluting stent (EES) has been developed as a second-generation DES. The study aim was to: (1) demonstrate that use of the EES in Japanese patients is non-inferior to use of the paclitaxel-eluting stent (PES) in US patients; and (2) compare vessel response to the EES in Japanese vs US patients. METHODS: The SPIRIT III Japan Registry, a prospective single-arm multicenter study was a part of the SPIRIT III global clinical program using harmonization by doing. The primary endpoint was in-segment late loss at 8 months, compared to US PES. RESULTS: A total of 88 subjects were enrolled in the Japan EES group. Angiographic in-segment late loss was significantly less in Japan EES vs US PES (0.15 ± 0.34 mm vs 0.28 ± 0.48 mm, respectively; P=.0185; Pnon-inferio r<.0001), while target vessel failure (TVF; 8.0% vs 9.9%) and major adverse cardiac events (MACE) at 9 months (5.7% vs 8.8%) were not significantly different between the 2 groups. No differences were observed between Japan and US EES populations in terms of late loss, TVF, or MACE. Neointimal volume and postprocedural incomplete stent apposition rate were lower in Japan EES vs US EES/PES. CONCLUSION: The SPIRIT III Japan Registry met the primary endpoint of lower late loss in the Japan EES group vs the US PES group, with comparable results for EES between the Japanese and US patients.