Effect of spacer diameter of the Dynesys dynamic stabilization system on the biomechanics of the lumbar spine: a finite element analysis.

2012 
Methods: In the finite element study, a previously validated lumbar spine model was used. Five Dynesys constructs with different spacer diameters (0.8, 0.9, 1.0, 1.1, and 1.2 times the original standard size) were implanted into the spine model and bore 4 displacement-controlled loading cases: flexion, extension, torsion, and lateral bending. Resultant range of motions (ROMs), disk stress, and facet contact forces at the bridged level and the cranial adjacent level were compared with the results of a spine model without Dynesys implantation.