Effect of Electrode Configuration and Capacitor Size on Internal Atrial Defibrillation Threshold Using Leads Currently Used for Ventricular Defibrillation.

1999 
Background: Previous studies have shown that endocardial atrial defibrillation, using lead configurations specifically designed for ventricular defibrillation, is feasible but the substantial patient discomfort might prevent the widespread use of the technique unless significant improvements in shock tolerability are achieved. It has been suggested that the peak voltage or the peak current but not the total energy delivered determines the patient pain perception and therefore, lower defibrillating voltage and current achieved with modifications in lead and waveforms may increase shock tolerability. This study was undertaken to evaluate the effect, on the atrial defibrillation threshold (ADFT), of the addition of a patch electrode (mimicking the can electrode) to the right ventricle (RV)-superior vena cava (SVC) lead configuration. The influence of capacitor size on ADFT using the RV-SVC+skin patch configuration was also assessed.