Direct enrichment of pathogens from physiological samples of high conductivity and viscosity using H-filter and positive dielectrophoresis

2018 
The full potential of microfluidic techniques as rapid and accurate methods for the detection of disease-causing agents and foodborne pathogens is critically limited by the complex sample preparation process, which commonly comprises the enrichment of bacterial cells to detectable levels. In this manuscript, we describe a microfluidic device which integrates H-filter desalination with positive dielectrophoresis (pDEP) for direct enrichment of bacterial cells from physiological samples of high conductivity and viscosity, such as cow's milk and whole human blood. The device contained a winding channel in which electrolytes in the samples continuously diffused into deionized (DI) water (desalination), while the bacterial cells remained in the samples. The length of the main channel was optimized by numerical simulation and experimentally evaluated by the diffusion of fluorescein into DI water. The effects of another three factors on H-filter desalination were also investigated, including (a) the flow rate ra...