Multicenter Evaluation of Edwards SAPIEN Positioning During Transcatheter Aortic Valve Implantation With Correlates for Device Movement During Final Deployment

2012 
Objectives This study sought to evaluate the exact location of Edwards SAPIEN (Edwards Lifesciences, Irvine, California) devices in different stages of implantation and to quantify possible operator-independent device movement during final deployment. Background Accurate device positioning during transcatheter aortic valve implantation is crucial in order to achieve optimal results. Methods This multicenter study consisted of 68 procedures with reliable pacemaker capture. Device positions were assessed using fluoroscopic images and the C-THV system (Paieon Medical, Rosh Ha'Ayin, Israel). Results The location after implantation was significantly higher than in the final stage of rapid pacing: 16.7 ± 16.3% of device height below the plane of the lower sinus border versus 32.6 ± 13.8%, p Conclusions The final Edwards SAPIEN position is mostly aortic in relation to the lower sinus border. There is an operator-independent upward movement of the device center during the final stage of implantation. Anticipated upward movement of the device should influence its positioning before final deployment.