Impact Fracture Resistance of Two Titanium‐Abutment Systems Versus a Single‐Piece Ceramic Implant
2011
Background: The number of patients with oral implants has increased significantly. However, the literature addressing the effect of impact force on titanium and/or ceramic implants is inconclusive. This study sought to determine the fracture resistance to impact load of titanium and ceramic endosseous oral implants.
Materials and Methods: Endosseous oral implants were vertically positioned in two different mounting media: brass and a bone-simulation material. The implant configurations tested included an experimental one-piece Y-TZP implant and a commercially available titanium implant (external hex) with both titanium and zirconia abutments. The specimens were subjected to an impact load using a pendulum impact tester with tup weights varying from 0.9 to 4.5 kg delivered at a radius of 40.64 mm. Loads were delivered to the abutment at a point 4.27 mm above the implant fixture and block junction. Statistical differences (p < .05) were established using the F-test for variances and, when different, t-test assuming unequal variances.
Results: For implants clamped in brass, the titanium implant with titanium abutment required the greatest energy to fracture the implant-abutment system (only the abutment screw failed). The ceramic implant and ceramic abutment on titanium implant presented the lowest fracture energy (p .25).
Conclusion: This investigation showed that the fracture energy of two titanium-abutment systems versus a single-piece Y-TZP implant in foam blocks simulating bone elastic modulus was not different, and that differences occurred when the embedding material elastic modulus was increased an order of magnitude.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
14
References
12
Citations
NaN
KQI