Contrast threshold of 4 full field digital mammography systems using different measurement methods

2006 
We compared three conspicuity tests applied to four full field digital mammography (FFDM) systems. The tests included: 1) the calculation of noise equivalent quanta (NEQ); 2) contrast-detail analysis with the CDMAM 3.4 phantom and 3) evaluation of the detectability of (simulated) microcal-cifications with specific well-known dimensions in mastectomy images. For each contrast-resolution test method, the exposure, processing and viewing conditions were identical. As a result, the only variable for a given test was the physical performance of the detector. The three test methods each rank the detectors in the same order. The flat-panel detector ranked the best overall, the dual-sided read-out storage phosphor detector ranked second and the single-sided-read-out storage phosphor detectors with 50 μm and 100 μm pixel sizes ranked similarly and were inferior to the other 2 detectors.