Performance of Single-Energy Metal Artifact Reduction in Cardiac Computed Tomography: A Clinical and Phantom Study

2020 
Abstract Background To investigate the performance of a reconstruction algorithm, single-energy metal artifact reduction (SEMAR), against standard reconstruction in cardiac computed tomography (CT) studies of patients with implanted metal and in a defibrillator lead phantom. Methods From a retrospective, cross-sectional clinical study with institutional review board approval of 118 patients with implanted metal, 122 cardiac CT studies from November 2009 to August 2016 performed on a 320-detector row scanner with standard and SEMAR reconstructions were included. The maximum beam hardening artifact radius, artifact attenuation variation surrounding the implanted metal, and image quality on a 4-point scale (1-no/minimal artifact to 4-severe artifact) were assessed for each reconstruction. A defibrillator lead phantom study was performed at different tube potentials and currents with both reconstruction methods. Maximum beam hardening artifact radius and average artifact attenuation variation were measured. Results In the clinical study, SEMAR markedly reduced the maximum beam hardening artifact radius by 77% (standard: 14.8 mm [IQR 9.7-22.2] vs. SEMAR: 3.4 mm [IQR 2.2-7.1], p Conclusions SEMAR reconstruction achieved superior image quality and markedly reduced maximum beam hardening artifact radius and artifact attenuation variation compared to standard reconstruction in 122 clinical cardiac CT studies of patients with implanted metal and in a defibrillator lead phantom study.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    16
    References
    4
    Citations
    NaN
    KQI
    []