Experimental validation of corrections factors for γ–γ and γ–X coincidence summing of 133Ba, 152Eu, and 125Sb in volume sources

2019 
True coincidence summing correction factors for 133Ba, 152Eu and 125Sb were determined experimentally for a small volume source and compared with correction factors obtained with three softwares (EFFTRAN-X, GESPECOR and VGSL). The radionuclides investigated have a relatively challenging decay scheme and their spectra are known to suffer from losses due to summation (γ–γ, γ–X and X–X) when measured at close distances on a HPGe detector sensitive to low energy photons. This study shows that the softwares were in good agreement with each other and the experimental data and the calculated activity was consistent with the activity in the volume source.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    14
    References
    2
    Citations
    NaN
    KQI
    []