The effects of pressure and flow on hemolysis caused by Bio-Medicus centrifugal pumps and roller pumps. Guidelines for choosing a blood pump.

1993 
Two Bio-Medicus BP-50 centrifugal pumps and two roller pumps were tested simultaneously with porcine blood at 21 o ±1 o C in four in vitro circuits to determine the effect of four combinations of flow and pressure conditions on blood damage. Flows of 300 ml/min (1/4-inch inner-diameter tubing in the roller pump) and 1775 ml/min (1/2-inch inner-diameter tubing in the roller pump) and pressure differences across the pump (ΔP=outlet pressure-inlet pressure) of 215 mm Hg (n=6) and 345 mm Hg (n=5) were examined. The index of hemolysis (milligrams plasma hemoglobin per 100 L blood pumped) for the BP-50 pump was higher at a flow of 300 ml/min than at a flow of 1775 ml/min (d<0.0002)