Prospective, randomized comparison of ProSealTM and ClassicTM laryngeal mask airways in anaesthetized neonates and infants
2009
Background When compared with the Classic™ laryngeal mask airway (cLMA), the recently introduced ProSeal™ laryngeal mask airway (PLMA) has modified features to produce higher airway seal pressures and enable ventilation in circumstances where the cLMA might fail. The first neonatal size 1 PLMA recently became available. This study was designed to compare the effectiveness of the size 1 cLMA and PLMA during positive pressure ventilation in anesthetized neonates and infants. Methods Forty-six consecutive patients undergoing elective cardiac surgical procedures were randomized for initial airway management with the cLMA or PLMA. Insertion time (IT), number of placement attempts, ease of placement, quality of the initial airway, maximum tidal volume (TV max ), and airway pressure at which an audible leak in the mouth ( P leak ) occurred were collected. All data were recorded before performing tracheal intubations. Results IT and success rate were similar for both LMAs. The initial quality of the airway was significantly better for the PLMA ( P max and P leak were significantly higher for PLMA (77 vs 58 ml, P vs 24.4 cm H 2 O, P Conclusions The size 1 PLMA forms a more effective seal than size 1 cLMA in neonates. This might allow the PLMA to be used in those newborn infants requiring high airway pressures for ventilation.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
23
References
20
Citations
NaN
KQI