Accuracy profile theory for the validation of an LC-MS-MS method for the determination of risperidone and 9-hydroxyrisperidone in human plasma.

2011 
A sensitive and specific LC-MS-MS method is described for the simultaneous quantification of risperidone and 9-hydroxyrisperidone in human plasma. After extraction with tert-butyl methyl ether, plasma samples were separated on an Atlantis HILIC Silica C18 column (4.6 × 150 mm, 5 µm) with a mobile phase of ammonium formate buffer (10 mM, pH 4.0)/acetonitrile (40/60, v/v). Detection was by MS-MS. The method was fully validated according to the accuracy profile theory. It is based on β-expectation tolerance interval for the total measurement error which includes trueness and intermediate precision. The measurement uncertainty derived from β-expectation tolerance interval was estimated at each of the validation standards. The linearity fitted well over the range of 0.11-26.75 ng mL–1 for risperidone with an LLOQ of 0.11 ng mL–1, and for 9-hydroxyrisperidone, at a range of 0.15-37.8 ng mL–1 with an LLOQ of 0.15 ng mL–1. The intra- and inter-batch precision of risperidone were <5.71 and 8.22%, respectively. For 9-hydroxyrisperidone, the data were 5.78 and 6.48%. The recoveries were 88.78% (risperidone) and 70.35% (9-hydroxyrisperidone). The developed method was applied to a pharmacokinetic study of risperidone.