Fast digital 4πβ−4πγ coincidence counting with offline analysis at IRA

2017 
Abstract IRA recently launched a project to digitize all the data acquisition systems it uses for primary radionuclide standardizations. It is well-known that the digital approach presents numerous advantages over the traditional analog electronics such as information losslessness, scalability, online and/or offline data processing, and it is also a solution to the growing difficulties to repair or renew ageing modules. As a first step in this wider program, our institute set-up a 4πβ−4πγ digital coincidence counting system, with FPGA (Field Programmable Gate Array)-based commercial boards from National Instruments (NI), to perform data acquisition and offline data analysis. Choosing all components and software from the same supplier provides a full compact and consistent electronic system. To demonstrate and validate the capacity of this system to standardize the activity of radioisotopes, we compare its predictions for the activity concentration of 133 Ba, 166 m Ho and 18 F solutions with the results from a coincidence counting system with analog electronics, as well as with the results from other primary methods and a secondary measurement performed with an IG11 ionization chamber (CIR, chambre d′ionization de reference ) with an equivalent activity traceable to the Systeme International de Reference .