New Interference Sensing Demand Pacemaker Functions

1978 
Demand cardiac pacemaker functions are under study that provide new methods to distinguish between cardiac activity and pulsatile electromagnetic interference (PEMI). All known forms of currently marketed ventricular inhibited demand pacemaker (VVI) functions can be inhibited by high level pulsatile electromagnetic interference. The recent introduction of shielded circuitry to protect against disruptive (inhibiting) EMI has reduced pacemaker sensitivity to interference. However, EMI received via the cardiac lead/electrode can still mimic cardiac activity. This mimicing occurs as a consequence of detection by defribrillator protection structures or by amplifier saturation from RF artifacts insufficiently suppressed by input QRS bandpass filters. The new functions under development employ a separate EMI detection receiver for controlling the pacemaker mode to minimize inhibition by PEMI.