Image artifact in dental cone-beam CT

2006 
Purpose The purpose of this study was to investigate the appearance and possible cause of an artifact seen in limited-volume cone-beam CT imaging. Methods A water-filled plastic cylinder was used as a phantom of the head. A test object was constructed as a bone-equivalent phantom to be imaged. The test object was variously positioned at the center of the phantom and near its margins. CT images of the test object were acquired using a 3DX Accuitomo system. Results In slice images with the test object positioned near the margin of the phantom, arch-shaped defects or deformities were observed on the side of the object. There was a negative correlation between the artifact and the CT value of the object. The artifact was larger in images scanned with a higher voltage. Conclusion The probability that this artifact is caused by halation from the image intensifier (II) system is suggested.