Setting pressure can change the size and shape of MRI artifacts caused by adjustable shunt valves: a study of the 4 newest models

2018 
OBJECTIVEAdjustable shunt valves that have been developed for the management of hydrocephalus all rely on intrinsically magnetic components, and artifacts with these valves on MRI are thus inevitable. The authors have previously reported that the shapes of shunt artifacts differ under different valve pressures with the proGAV 2.0 valve. In the present study the authors compared the size and shape of artifacts at different pressure settings with 4 new-model shunt valves.METHODSThe authors attached 4 new models of MRI-resistant shunt valve to the temporal scalp of a healthy volunteer: the proGAV 2.0; Codman Certas Plus; Polaris; and Strata MR. They set 3 different scales of pressures for each valve, depending on magnet orientation to the body axis. Artifacts were evaluated and compared among all valves on a 3.0-T GE scanner and 2 valves were also evaluated on a Philips scanner and a Siemens scanner. In-plane artifact sizes were evaluated as the maximum distance of the artifact from the expected scalp.RESULT...