30-ps time resolution with segmented scintillation counter for MEG II

2017 
Abstract A new timing detector has been developed to measure ∼50 MeV/c positrons with a time resolution of σ t ≃ 30 ps in the MEG II experiment. The detector are segmented into 512 scintillation counters, each of which consists of 120 × ( 40 or 50 ) × 5 mm 3 size BC-422 and two arrays of six AdvanSiD silicon photomultipliers. The single-counter resolutions are measured to be 70–80 ps. The counter layout is optimized to get the maximum number of hit counters (on average 9 for signal positrons). This multiple-counters measurement leads to a significant improvement in the time resolution down to 30 ps. Using the first one-fourth (128) counters, a pilot run was carried out using the MEG II beam of 7 × 10 7 μ + / s and the basic functionality was tested.