Lesion Index Titration Using Contact-Force Technology Enables Safe and Effective Radiofrequency Lesion Creation at the Root of the Aorta and Pulmonary Artery

2019 
Background: Ablation of some myocardial substrates requires catheter-based radiofrequency delivery at the root of a great artery. We studied the safety and efficacy parameters associated with catheter-based radiofrequency delivery at the root of the aorta and pulmonary artery. Methods: Thirty-six pigs underwent in-vivo catheter-based ablation under continuous contact-force and lesion index (power, contact-force, and time) monitoring during 60-s radiofrequency delivery with an open-irrigated tip catheter. Twenty-eight animals were allocated to groups receiving 40 W (n=9), 50 W (n=10), or 60 W (n=9) radiofrequency energy, and acute (n=22) and chronic (n=6) arterial wall damage was quantified by multiphoton microscopy in ex vivo samples. Adjacent myocardial lesions were quantified in parallel samples. The remaining 8 pigs were used to validate safety and efficacy parameters. Results: Acute collagen and elastin alterations were significantly associated with radiofrequency power, although chronic assessment re...