Characteristics of Un doped and Europium-dopedSrI2 Scintillator Detectors

2012 
High energy resolution gamma-ray detectors that can be formed into relatively large sizes while operating at room temperature offer many advantages for national security applications. We are working toward that goal through the development of SrI{sub 2}(Eu) scintillator detectors, which routinely provide ;10 cm{sup 3}. In this study, we have tested pure, undoped SrI{sub 2} to gain a better understanding of the scintillation properties and spectroscopic performance achievable without activation. An undoped crystal grown from 99.999% pure SrI{sub 2} pellets was tested for its spectroscopic performance, its light yield, and uniformity of scintillation light collection as a function of gamma-ray interaction position relative to the crystal growth direction. Undoped SrI{sub 2} was found to provide energy resolution of 5.3% at 662 keV, and the light collection nonuniformity varied by only 0.72% over the length of the crystal. Measurements of both a 3% Eu-doped and the undoped SrI{sub 2} crystal were carried out in the SLYNCI facility and indicate differences in their light yield non-proportionality. The surprisingly good scintillation properties of the pure SrI{sub 2} crystal suggests that with high-purity feedstock, further reduction of the Eu concentration can be made to grow largermore » crystals while not adversely impacting the spectroscopic performance.« less