Noninvasive investigation of exocrine pancreatic function: Feasibility of cine dynamic MRCP with a spatially selective inversion-recovery pulse.
2015
Purpose
To investigate the feasibility of noncontrast-enhanced cine dynamic magnetic resonance cholangiopancreatography (MRCP) with a spatially selective inversion-recovery (IR) pulse for evaluating exocrine pancreatic function in comparison with the N-benzoyl-L-tyrosyl-p-aminobenzoic acid (BT-PABA) test as a pancreatic exocrine function test.
Materials and Methods
Twenty subjects with or without chronic pancreatitis were included. MRCP with a spatially selective IR pulse was repeated every 15 seconds for 5 minutes to acquire a total of 20 images (cine-dynamic MRCP). The median and mean frequency of the observation (the number of times) and the moving distance (mean secretion grading scores) of pancreatic juice inflow on cine-dynamic MRCP were compared with a BT-PABA test.
Results
The urinary PABA excretion rate (%) had significant positive correlations with both the mean secretion grade (r = 0.66, P = 0.002) and frequency of secretory inflow (r = 0.62, P = 0.004) in cine dynamic MRCP. Both the mean frequency of observations of pancreatic secretory inflow (1.4 ± 1.6 times vs. 14.3 ± 4.2 times, P < 0.001) and the mean secretion grade (grade = 0.16 ± 0.24 vs. grade = 1.81 ± 0.81, P < 0.001) was significantly lower in the chronic pancreatitis group than in the normal subject group.
Conclusion
Cine dynamic MRCP with a spatially selective IR pulse may have potential for estimating the pancreatic exocrine function noninvasively as a substitute for the BT-PABA test. J. Magn. Reson. Imaging 2015;42:1266–1271.