[Determination of p-Anisidine in workplace air by HPLC].
2018
Objective
To establish an HPLC method for determining p-Anisidine in workplace air.
Methods
An C18 capillary column was used with the detection wavelength of 240 nm, the methanol solution (v/v=62:38) as mobile phase, the flow rate was 0.9 ml/min, and the injection volume was 20 μl.
Results
The linear range (μg/ml) for p-Anisidine was 0~100; correlation coefficient was 0.999 7; the detection limit was 0.15 μg/ml; the average precision of RSD was 4.4%. The sampling efficiency was more than 95% using silicone tube as sampling filter. Sample can be stored at room temperature for more than 7 d.
Conclusion
the method appered to be accurate, simple, sensitive and reliable, also have a high sampling efficiency, and can be usd to determine p-anisidine in workplace air simultaneously.
Key words:
chromatography; High performance liquid; Workplace; P-Anisidine