Detecting High Hyperopia: The Plus Lens Test and the Spot Vision Screener
2017
PURPOSE: To evaluate the usefulness of the Plus Lens (Goodlite Company, Elgin, IL) test and the Spot Vision Screener (Welch Allyn, Skaneateles Falls, NY) in detecting high hyperopia in a pediatric population. METHODS: Between June and August 2015, patients were screened with the Spot Vision Screener and the Plus Lens test prior to a scheduled pediatric ophthalmology visit. The following data were analyzed: demographic data, Plus Lens result, Spot Vision Screener result, cycloplegic refraction, and examination findings. Sensitivity/specificity and positive/negative predictive values were calculated for the Plus Lens test and Spot Vision Screener in detecting hyperopia as determined by the "gold-standard" cycloplegic refraction. RESULTS: A total of 109 children (average age: 82 months) were included. Compared to the ophthalmologist's cycloplegic refraction, the Spot Vision Screener sensitivity for +3.50 diopters (D) hyperopia was 31.25% and the specificity was 100%. The Plus Lens sensitivity for +3.50 D hyperopia was 43.75% and the specificity was 89.25%. Spot Vision Screener sensitivity increased with higher degrees of hyperopia. CONCLUSIONS: In this preliminary study, the Plus Lens test and the Spot Vision Screener demonstrated moderate sensitivity with good specificity in detecting high hyperopia. [J Pediatr Ophthalmol Strabismus. 2017;54(3):163-167.].