Free release static coating of glass capillary columns; Static coating at elevated pressure — Theoretical considerations and practice

1984 
Static coating of glass capillary columns has hitherto solely been carried out by evaporation of the stationary phase solvents under vacuum conditions. However, since a solvent vapour pressure higher than the external pressure is the only requirement for the vapour to flow out of the column, evacuation should not be necessary. Several important factors in the static coating procedure, such as mass (solvent vapour) — and heat transfer, heat of vaporisation of the solvent and viscosity of the stationary phase solution at elevated temperature and outlet pressure are discussed, principally to rationalise an improved static coating procedure. The alternative, so-called free-release static coating procedure, was evaluated practically by coating several columns with OV-101 and SE-30. Coating speed was found to be rapid and relatively constant whereas coating efficiency was between 80 and 100%.