Dialysis membrane-dependent removal of middle molecules during hemodiafiltration: the beta2-microglobulin/albumin relationship.

2004 
Aim: Current hemodialysis therapy modalities such as online hemodiafiltration (HDF) attempt to enhance solute removal over a wide molecular weight range through a combination of diffusion and convection. While the effects of variations of treatment modalities and conditions have been studied reasonably well, few studies have examined the efficacy of HDF to remove middle molecules in relation to the dialyzer and membrane characteristics. In this investigation, diverse high-flux dialyzers, covering a wide range of membrane permeabilities, were compared under identical in vivo conditions to assess their ability to eliminate larger uremic retention solutes (using β 2 -microglobulin as a surrogate of middle molecules) without simultaneously causing excessive leakage of useful proteins such as albumin. Patients and methods: In a prospective, crossover study, 3 ESRD patients were treated with 8 different brands of high-flux dialyzers at 4 different ultrafiltration (UF)/ substitution flow rates (QS: 0, 30, 60, 90 ml/ min) in post-dilution HDF mode. Thus, each patient underwent 32 treatment sessions, with a total of 96 treatment sessions conducted during the entire clinical study. Albumin and β 2 -microglobulin levels were measured in both, dialysate and blood. Both, albumin and β 2 -microglobulin elimination was dependent upon the permeability of the dialysis membrane as well as on the ultrafiltration/substitution flow rates applied. Results: At the maximum UF rate of 90 ml/min, the total albumin loss (measured in the dialysate) ranged from 300 mg/4 h (for the FLX-15 GWS dialyzers) to 7,000 mg/4 h (for the BS-1.3U dialyzers). Up to 50% reduction of albumin occurred within the first 30 minutes of the dialysis treatment, and the leakage of albumin increased exponentially with increasing UF rates as well as increasing transmembrane pressure (TMP). The various dialyzers could be classified according to their UFR-dependent β 2 -m reduction rates (RR), into low ( 70%; BS-1.3U, APS 650, FX 60) removers of middle molecules. One dialyzer type (CT 150G) showed extremely low β 2 -m RR and relatively high albumin losses. Most membranes, however, showed either low albumin leakage coupled with low β 2 -m removal, or high β 2 -m RR but at the expense of considerable albumin leakage. Only 2 membrane types approached the desired balance between high to medium β 2 -m RR while simultaneously restricting the albumin leakage especially at higher filtration/substitution rates. Conclusion: Our investigations demonstrate that not all dialysis membranes classified as high-flux are comparable in their ability to specifically and efficiently remove middle molecules, or curtail the unwanted excessive leakage of essential proteins from the patient's blood. Thus, the selection of appropriate high-flux dialyzers for specific patient requirements should be based more upon clinical evaluations and analyses rather than on product specifications alone.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    54
    Citations
    NaN
    KQI
    []