Optimization of scan timing for aortic computed tomographic angiography using the test bolus injection technique

2016 
BackgroundWith fast computed tomography (CT), it is possible for the scanning to outpace the contrast medium bolus during aortic CT angiography (CTA).PurposeTo evaluate the effectiveness of a new method for reducing the risk of outpacing in which the scan start timing (ST) and speed can be estimated from the peak enhancement time measured at the femoral artery using a single test-bolus injection (femoral artery test injection method [FTI method]).Material and MethodsIn 30 cases of aortic CTA, we measured the time to peak enhancement at the femoral artery (TPF) and the ascending aorta (TPA) with test-bolus injection performed twice in each examination. From the resultant linear relationship between TPF and transit time (TT = TPF − TPA), we developed a method for determining the ST and TT from TPF. One hundred patients were assigned to two groups: FTI and bolus tracking (BT), each with 50 patients. CT values were measured in main vessels (ascending aorta, descending aorta, femoral artery). The CT values of ...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    24
    References
    2
    Citations
    NaN
    KQI
    []