Can CT scan protocols used for radiotherapy treatment planning be adjusted to optimize image quality and patient dose? A systematic review

2017 
This article reviews publications related to the use of CT scans for radiotherapy treatment planning, specifically the impact of scan protocol changes on CT number and treatment planning dosimetry and on CT image quality. A search on PubMed and EMBASE and a subsequent review of references yielded 53 relevant articles. CT scan parameters significantly affect image quality. Some will also affect Hounsfield unit (HU) values, though this is not comprehensively reported on. Changes in tube kilovoltage and, on some scanners, field of view and reconstruction algorithms have been found to produce notable HU changes. The degree of HU change which can be tolerated without changing planning dose by >1% depends on the body region and size, planning algorithms, treatment beam energy and type of plan. A change in soft-tissue HU value has a greater impact than changes in HU for bone and air. The use of anthropomorphic phantoms is recommended when assessing HU changes. There is limited published work on CT scan protocol ...