Image guidance based on MRI for spinal interstitial laser thermotherapy: technical aspects and accuracy

2017 
OBJECTIVEImage guidance for spinal procedures is based on 3D-fluoroscopy or CT, which provide poor visualization of soft tissues, including the spinal cord. To overcome this limitation, the authors developed a method to register intraoperative MRI (iMRI) of the spine into a neuronavigation system, allowing excellent visualization of the spinal cord. This novel technique improved the accuracy in the deployment of laser interstitial thermal therapy probes for the treatment of metastatic spinal cord compression.METHODSPatients were positioned prone on the MRI table under general anesthesia. Fiducial markers were applied on the skin of the back, and a plastic cradle was used to support the MRI coil. T2-weighted MRI sequences of the region of interest were exported to a standard navigation system. A reference array was sutured to the skin, and surface matching of the fiducial markers was performed. A navigated Jamshidi needle was advanced until contact was made with the dorsal elements; its position was confir...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    25
    References
    23
    Citations
    NaN
    KQI
    []