Measurement of radionuclides using ion chromatography and flow-cell scintillation counting with pulse shape discrimination
1996
Principal conclusions are: CsI(Tl) provides sufficient pulse shape discrimination for use in the flow-cell detector. However, an improved method of coating is needed to extend the useful life of a detection cell. Of the activation/fission products investigated, only the co-elution of {sup 137}Cs and {sup 63}Ni produced a radiological interference. Tritium (and presumably other non-ionic radioisotopes) can be separated during the loading of the solution onto the pre- concentration column. Natural U (and/or decay products) produced a radiological interference with {sup 90}Sr. This is a potential problem. No potential radiological interferences were observed with {sup 223}Th. Chemical interferences were observed to some degree for all the chemicals tested except for the chloride solutions, NaCl and KCl, and the sulfate solution, Na{sub 2}SO{sub 4}. The specific interference effects were decreased detection efficiencies and changes in peak elution times. The NEL`s (non-observable effects loadings) are tentative targets for development of sample processing protocols, which is the next phase of the work.