Multicentre evaluation of a two-step automated enzyme immunoassay of free thyroxine

1992 
: The two-step enzymatic immunoassay of free thyroxine (IMx FT4, Abbott Laboratories, Chicago, IL) was studied in three centres. The assay involved a fluorimetric measurement and took 45 minutes using a completely automated procedure. The results were compared with those from the free thyroxine two-step radioimmunoassay and with the "calculation" of free thyroxine. The analytical precision was found to be excellent if the analyser was correctly set. The IMx FT4 assay seemed unaffected by increased concentrations of albumin and of non-esterified fatty acids (oleic acid) up to 5 mmol/l. The euthyroid reference interval, defined as that including 95% of 194 control subjects, was 12-21 pmol/l. A limited overlap existed between euthyroid and hyperthyroid patients, but a larger one was seen between the euthyroid and hypothyroid population, the latter including subclinical hypothyroidism. IMx FT4 results agreed well when compared with those from two-step radioimmunoassays. The IMx FT4 technique gave rise to a low percentage of elevated results in patients being treated with heparin, but was undisturbed by autoantibodies to thyroxine and triiodothyronine which were present in one hypothyroid patient.