Development of the permeability/performance reference compound approach for in situ calibration of semipermeable membrane devices

2002 
Permeability/performance reference compounds (PRCs) are analytically noninterfering organic compounds with moderate to high fugacity from semipermeable membrane devices (SPMDs) that are added to the lipid prior to membrane enclosure. Assuming that isotropic exchange kinetics (IEK) apply and that SPMD−water partition coefficients are known, measurement of PRC dissipation rate constants during SPMD field exposures and laboratory calibration studies permits the calculation of an exposure adjustment factor (EAF). In theory, PRC-derived EAF ratios reflect changes in SPMD sampling rates (relative to laboratory data) due to differences in exposure temperature, membrane biofouling, and flow velocity−turbulence at the membrane surface. Thus, the PRC approach should allow for more accurate estimates of target solute/vapor concentrations in an exposure medium. Under some exposure conditions, the impact of environmental variables on SPMD sampling rates may approach an order of magnitude. The results of this study sug...