Preliminary outcomes of the cementless UNITED hip system for primary total hip arthroplasty at a minimum 2-year follow-up

2021 
This study aimed to report the initial results of the cementless UNITED hip system in primary total hip arthroplasty (THA) with a minimum follow-up of 2 years. We retrospectively studied a consecutive series of 203 cementless THAs in 180 patients operated between 2015–2017. We included 89 female and 91 male patients with a mean age of 67 (28 to 89) years. The mean follow-up was 40 (29 to 62) months. Clinical outcome scores and radiographs were measured. Survival was calculated defining failure as the need for any further femoral or acetabular revision, irrespective of the reason. No femoral component loosening was detected. One patient had a Vancouver-B1 intraoperative periprosthetic femoral fracture treated with implant retention and cerclage wires. Two acetabular components were revised for aseptic loosening. Three patients suffered an acute infection treated with debridement, antibiotics, and implant retention. The mean Merle d'Aubigne et Postel scores improved from 13 (4 to 16) points preoperatively to 17 (12 to 18) points at the latest follow-up (p < 0.001). At a mean time of 40 months of follow-up, the survival was 99% and 100% for the acetabular and the femoral components, respectively. This cementless design showed excellent preliminary outcomes in terms of fixation and patient satisfaction, comparable to that of other well-known similar systems.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    27
    References
    0
    Citations
    NaN
    KQI
    []