Fluorescent Detection of Oral Pathogens by a Solid-Phase Immunoassay on PDMS

2005 
We have developed an array of sensors for the oral pathogen Streptococcus mutans (S. mutans) using an enzyme-linked immunosorbent assay (ELISA) on a polydimethylsiloxane (PDMS) device. The model bacterial analyte, S. mutans, has been implicated in the initiation and progression of dental caries. The PDMS was modified with 3-aminopropyltriethoxysilane (APTS) and glutaraldehyde to covalently crosslink monoclonal anti-S. mutans immunoglobulin G (IgG) to the sensor surface. Successful IgG immobilization was verified by AFM and fluorescence imaging. Colloidal bacteria were captured on the sensor surface and labeled with immuno-active quantum dots (QDs), whose fluorescence was excited by an LED and detected by a CCD. The system was capable of detecting S. mutans concentrations as low as 6times10 6 cells/ml in a 20 mul sample. This work represents a stable foundation for the development of a chair side diagnostic system capable of specific and sensitive detection of pathogens directly from oral fluid
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    5
    References
    3
    Citations
    NaN
    KQI
    []