Echocardiographic Ramp Test for Continuous-Flow Left Ventricular Assist Devices : Do Loading Conditions Matter?

2015 
Abstract Objectives This study investigated whether continuous AI and/or elevated mean arterial pressure (MAP) were associated with false positive results for flow obstruction in echocardiographic ramp speed tests in patients with a continuous-flow left ventricular assist device. Background Failure to reduce the left ventricular end-diastolic diameter (LVEDD) with increasing device speeds in a ramp test is predictive of pump obstruction. Aortic insufficiency (AI) or increased MAP can diminish the ability to unload the left ventricle. Methods LVEDD was plotted against device speed, and a linear function slope was calculated. A flat LVEDD slope (≥–0.16) was considered abnormal (suggestive of obstruction). Ramp test results were compared in patients with or without either AI or increased MAP at baseline speed, and receiver-operator characteristic (ROC) curves were constructed for predictors of device obstruction. Device thrombosis was confirmed by direct visualization of clot at explantation or on inspection by the manufacturer. Results Of 78 ramp tests (55 patients), 36 were abnormal (18 true positive, 18 false positive), and 42 were normal (37 true negative, 5 false negative). In patients with AI, LVEDD slope was –0.14 ± 0.17, which was consistent with device obstruction (vs. –0.25 ± 0.11 in patients without AI; p  Conclusions Abnormal loading conditions due to AI or elevated MAP may result in false positive ramp tests.