The effective dose assessment of C-arm CT in hepatic arterial embolisation therapy

2013 
Objective:To assess the effective dose of the liver C-arm computed tomography (CT) scan during hepatic arterial embolisation surgery with clinical dose–area product (DAP) data from Taiwan.Methods:The experiment used two kinds of phantoms: RANDO® Man and RANDO Woman (The Phantom Laboratory, Salem, NY), embedded with thermoluminescent dosemeters at locations according to the International Commission on Radiological Protection 103 report. The conversion factors of DAP to effective doses for males and females, respectively, were obtained. The clinical DAP data of liver C-arm CT scan during hepatic arterial embolisation surgery were collected in a hospital in Taiwan.Results:There were 125 liver transarterial embolisation therapy cases, including 94 males and 31 females, from February 2009 to June 2010. C-arm CT was used 38 times for males and 17 times for females. The corresponding average and standard deviation of clinical DAP were 61.0±6.6 Gy cm2 and 52.2±8.3 Gy cm2, respectively.Conclusion:The DAP of RANDO ...