Development and validation of sensitive LC/MS/MS method for quantitative bioanalysis of levonorgestrel in rat plasma and application to pharmacokinetics study

2015 
Abstract Rapid, sensitive, selective and accurate LC/MS/MS method was developed for quantitative determination of levonorgestrel (LNG) in rat plasma and further validated for specificity, linearity, accuracy, precision, sensitivity, matrix effect, recovery efficiency and stability. Liquid–liquid extraction procedure using hexane:ethyl acetate mixture at 80:20 v:v ratio was employed to efficiently extract LNG from rat plasma. Reversed phase Luna column C18(2) (50 × 2.0 mm i.d., 3 μM) installed on a AB SCIEX Triple Quad™ 4500 LC/MS/MS system was used to perform chromatographic separation. LNG was identified within 2 min with high specificity. Linear calibration curve was drawn within 0.5–50 ng·mL −1 concentration range. The developed method was validated for intra-day and inter-day accuracy and precision whose values fell in the acceptable limits. Matrix effect was found to be minimal. Recovery efficiency at three quality control (QC) concentrations 0.5 (low), 5 (medium) and 50 (high) ng·mL −1 was found to be >90%. Stability of LNG at various stages of experiment including storage, extraction and analysis was evaluated using QC samples, and the results showed that LNG was stable at all the conditions. This validated method was successfully used to study the pharmacokinetics of LNG in rats after SubQ injection, providing its applicability in relevant preclinical studies.