Daily Biopsy Diagnosis in Surgical PathologyConcordance Between Light Microscopy and Whole-Slide Imaging in Real-Life Conditions

2018 
Objectives: The current challenge for the various digital whole-slide imaging (WSI) systems is to be definitively validated for diagnostic purposes. We designed a concordance study between glass slide and digital slide diagnosis in real-life conditions, coupled with an ergonomic study. Methods: Three senior pathologists evaluated, first in glass slides and then in digital slides, 119 biopsy cases, including 749 slides, with 332 H&E saffron stains and 417 additional techniques, mainly immunohistochemistry. Results: All digital slides, including specially stained slides, were interpretable. Concordance between glass slides and digital slides was observed in 87.4% of cases. Minor discordances were observed in 12 (10.1%) cases and major discordances, with therapeutic impact, in three (2.5%), including one related to WSI. The satisfaction of participants was high and increased with time. Conclusions: Our study confirms the feasibility and accuracy of WSI diagnosis, even for cases having multiple samples and requiring special staining techniques, such as immunohistochemistry and in situ hybridization.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    38
    References
    7
    Citations
    NaN
    KQI
    []