Remote Monitoring Reduces Healthcare Use and Improves Quality of Care in Heart Failure Patients With Implantable DefibrillatorsClinical Perspective

2012 
Background— Heart failure patients with implantable cardioverter-defibrillators (ICDs) or an ICD for resynchronization therapy often visit the hospital for unscheduled examinations, placing a great burden on healthcare providers. We hypothesized that Internet-based remote interrogation systems could reduce emergency healthcare visits. Methods and Results— This multicenter randomized trial involving 200 patients compared remote monitoring with standard patient management consisting of scheduled visits and patient response to audible ICD alerts. The primary end point was the rate of emergency department or urgent in-office visits for heart failure, arrhythmias, or ICD-related events. Over 16 months, such visits were 35% less frequent in the remote arm (75 versus 117; incidence density, 0.59 versus 0.93 events per year; P =0.005). A 21% difference was observed in the rates of total healthcare visits for heart failure, arrhythmias, or ICD-related events (4.40 versus 5.74 events per year; P <0.001). The time from an ICD alert condition to review of the data was reduced from 24.8 days in the standard arm to 1.4 days in the remote arm ( P <0.001). The patients' clinical status, as measured by the Clinical Composite Score, was similar in the 2 groups, whereas a more favorable change in quality of life (Minnesota Living With Heart Failure Questionnaire) was observed from the baseline to the 16th month in the remote arm ( P =0.026). Conclusions— Remote monitoring reduces emergency department/urgent in-office visits and, in general, total healthcare use in patients with ICD or defibrillators for resynchronization therapy. Compared with standard follow-up through in-office visits and audible ICD alerts, remote monitoring results in increased efficiency for healthcare providers and improved quality of care for patients. Clinical Trial Registration— URL: . Unique identifier: [NCT00873899][1]. # Clinical Perspective {#article-title-37} [1]: /lookup/external-ref?link_type=CLINTRIALGOV&access_num=NCT00873899&atom=%2Fcirculationaha%2F125%2F24%2F2985.atom
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    38
    References
    0
    Citations
    NaN
    KQI
    []