The impact of eculizumab on routine complement assays

2018 
Abstract Background Eculizumab (ECU) blocks complement C5 cleavage, preventing the formation of C5a and the cytolytic effects of the membrane attack complex. The presence of ECU in blood impacts routine complement tests used to monitor treatment. Methods Residual serum samples with normal total complement (CH50) and residual citrate plasma with normal PT/APTT were spiked with ECU at varied concentrations ranging from 25 to 600 μg/mL. In addition, seventy-one samples from patients on ECU were obtained. Artificial and patient samples were analyzed for CH50 and C5 function (Wako Diagnostics), C5 concentration (Quidel), AH50 (Wieslab ELISA) and sMAC (Quidel). ECU concentration was measured by mass spectrometry for all patients. Results Complement blockage by ECU was evident in spiked artificial samples. At 25 μg/mL ECU, partial complement blockage was observed in CH50, AH50 and C5 function in serum. Complete blockage defined by undetectable AH50 ( 100 μg/mL, C5 function was Conclusions The spiked sera and patient samples showed complement blockage with CH50, AH50 and C5 function assays when ECU >100 μg/mL. CH50, AH50 or C5 function assays can serve as indicators for the pharmacodynamic effects of eculizumab . Allied to ECU concentration, laboratory studies may be helpful to tailor therapy.